FAQ

R-Motion

1 - Which parameters should be applied to R-Motion® (angles & speed)?

R-Motion® files have been designed and can be used with standard reciprocating modes in existing motors (CCW >> CW).
FKG is launching a cordless endomotor with a preset program dedicated to R-Motion® files and an integrated apex locator: the Rooter® Universal
For more information about this new motor, please contact your local FKG dealer or consult the following link: https://www.fkg.ch/products/endodontics/motor-apex-locator/rooter-universal

 

2 - Can I use R-Motion® files on OTR (Optimum Torque Reverse) motion?

OTR motion is not recommended by FKG for the use of R-Motion® files.

 

3 - What is the metallurgy of R-Motion® and how does it affect the treatment?

All R-Motion® instruments are made of a medical grade NiTi alloy. The active part of the instrument undergoes a proprietary heat treatment that triggers a phase transition (between martensite and austenite) just below body temperature, between 32° and 35°C.

 

4 - Does R-Motion® have constant or variable taper ?

  • R-Motion® Glider has a taper of 3% from the tip to 9mm, then 6%.
  • R-Motion® shaping files have a constant taper.

 

5 - Which cross section for R-Motion®?

Rounded triangular symmetrical cross-section in all instruments.

 

6 - Is R-Motion® recommended for retreatment?

R-Motion® has not been validated for retreatment cases and therefore cannot be recommended for this indication.
FKG recommends using its existing dedicated systems: XP-endo Retreatment or D-Race.

 

 

RACE EVO

1 - What is the metallurgy of RACE® EVO and how does it affect the treatment?

All RACE® EVO instruments are made of a medical grade NiTi alloy.The active part of the instrument undergoes a proprietary heat treatment that triggers a phase transition (between martensite and austenite) just below body temperature, between 32°and 35°C.

 

2 - What is the impact of this new metallurgy on the cutting efficiency of RACE® EVO 4% sequence and RACE® EVO 6% sequence instruments?

The heat treatment does not have a direct influence on the cutting efficiency, it is rather the legacy RaCe design associated with the higher rotation speed (1’000 rpm) that improves the cutting efficiency, by up to 50% compared to former RaCe product lines.

 

3 - What are the resistance to cyclic fatigue and torsional resistance of RACE® EVO 4% sequence and RACE® EVO 6% sequence instruments?

RACE® EVO 4% sequence and RACE® EVO 6% sequence instruments show up to 2.8 times higher resistance to cyclic fatigue and up to 1.9 times higher resistance to torsion.

 

4 - Why is RACE® EVO designed with a safety tip rather than with a booster or active tip?

RACE® EVO’s unique rounded safety tip allows for an efficient progression of the instrument along the canal.

 

5 - Why is it not necessary to have an intermediate file between the rotary glide path file (15/.04) and the initial shaping file (25/.04)?

Thanks to the combination of heat treatment, design and higher speed, an intermediate file is not necessary. You will be convinced when you try it !

 

6 - How many root canals can I treat with a RACE® EVO instrument?

With any RACE® EVO instrument you can treat up to a maximum of 8 root canals. However, you still need to inspect the RACE® EVO instrument on a regular basis during use, and change if it shows any signs of deformation (for example: straightening). The good news is that RACE® EVO instruments, once they have reached the end of their service life, will deform in a visible way before they fracture. This warning is an additional safety feature specific to these instruments.

 

7 - Without SMD (SafetyMemoDisc), how can I know how many canals I treated?

SMD is not needed because RACE® EVO instruments show sufficient resistance to fatigue to guarantee a maximum use of 8 canals, no matter how difficult the curvatures are. However, if you feel more confident with SMD, they are still available separately in the FKG product assortment.

 

8 - Is there any risk if I use RACE® EVO instruments over 1’000 rpm?

To achieve full efficiency of RACE® EVO instruments, FKG recommends a speed of use between 800 and (ideally) 1'000 rpm. FKG cannot recommend the use of RACE® EVO instruments at higher or lower speeds.

 

9 - Is RACE® EVO recommended for retreatment?

RACE® EVO has not been validated for retreatment cases and therefore cannot be recommended for this indication. FKG recommends using its existing dedicated systems: XP-endo Retreatment or D-Race.

 

 

Drehzahl & Drehmoment - Empfehlungen

CI-Metallschaft für Winkelstück - Farbencodes

Die rotierenden Instrumente von FKG sind mit CI-Metallschaft, Längenmarkierung und Silikon-Endo-Stops ausgestattet.

Damit soll eine einfache Identifikation von ISO-Durchmesser (breiter Ring) und Konizität (schmaler Ring) erreicht werden. Die Informationen bleiben erkennbar, wenn das Instrument in den Kopf des Winkelstücks eingesteckt wird.

 

Farbencodes

 

CI-Metallschäfte für Winkelstück - Längenmarkierung

Die Längenmarkierungen dienen zur Kennzeichnung der Arbeitslänge im Zahn, sie ergänzen den Endo stop. Die Längenmarkierungen werden auf allen Instrumenten der Produktfamilien XP-endo und Race aufgebracht.

  • Längenmarkierungen aufgebracht auf Instrumenten der Länge 21 mm/25 mm/31 mm

 

Beispiele für Instrumente mit Längenmarkierungen :

 

CM-Metallschäfte für Winkelstück

Der Farbring am Schaft gibt die ISO-Durchmesser an, die SafetyMemoDisc-Scheibe (SMD) identifiziert die Konizität der Instrumente.

 

Silikon Endo stop

Zur Markierung der Arbeitslänge, röntgenopak.

  • Die Markierung zeigt die ursprüngliche Position des Intruments im Kanal an (Stahlinstrumente).

  • ISO-Farben geben die Feilenlänge an.

 

Race - Anwendervorteile

Alle Race-Instrumente haben einzigartige Eigenschaften, welche die Anwendungssicherheit erhöhen:

  • Einzigartige abgerundete Sicherheitsspitze: Präzise Führung, gleitet über Unebenheiten hinweg.
  • Patentiertes Design mit alternierenden Schneidkanten: Praktisch kein Einschraub-Effekt.
  • Scharfe Schneidekante: Optimale Schneideffizienz, dadurch Zeitgewinn.
  • Spezielle elektrochemische Politur: Erhöhte Widerstandsfähigkeit gegen Torsion und zyklische Ermüdung.
  • SafetyMemoDisc (SMD): Kontrolle der Metallermüdung und der Zahl der Anwendungen.

Race - Goldene Regeln zur optimalen Anwendung der Race-Instrumente

  • Drehzahl: 600 - 1000/min - Drehmoment: 0.5 - 1.5 Ncm, je nach Instrumente Details sehen.
  • Weite Auf-und-Ab-Bewegungen, ohne jemals zu viel Druck auszuüben.
  • Leichte Hand, das Instrument arbeiten lassen.
  • Nur drei bis vier Sekunden am Stück arbeiten, dann herausziehen.
  • Schneide reinigen und Kanal spülen.

Race - Mit welchen Motoren können die Instrumente der Race-Produktfamilie eingesetzt werden?

Motoren der Dentaleinheit, um ca. 600 U/ min zu erreichen, verwenden Sie für:

  • Luftmotoren mit Drehzahl 20.000 U/min, ein Untersetzungswinkelstück 32:1.
  • Elektrische Motoren mit Drehzahl 40.000 U/min, Untersetzungswinkelstück 64:1, 70:1.

Endodontie-Motoren, kabelgebunden, oder kabellos, wie der Rooter:

  • Einstellung der Drehzahl auf 600-1000 U/min und des Drehmoments auf 0.5-1.5 Ncm.

Race - Wie oft kann man ein Race-Instrument verwenden?

Nachfolgend die mittels der SafetyMemoDiscs gegebenen Indikationen:

  • 1 Segment bei leichten Fällen (S), geraden, leicht gekrümmten oder weiten Kanälen.
  • 2 Segmente bei mittelschweren Fällen (M), engeren oder stark gekrümmten Kanälen.
  • 4 Segmente bei schwierigen Fällen (D), sehr stark gekrümmten, S-förmigen, sehr engen oder kalzifizierten Kanälen.

Auch wenn die Race-Instrumente sterilisierbar und mehrfach wiederverwendbar sind, ist ihre Verwendung nach dem Prinzip «nur für einen Patienten» empfehlenswert, um jegliches Risiko einer Kreuzkontamination auszuschließen.

TotalFill info collection

Mehr Info hier